Options
A clinical study to validate the pupil rescaling technique by using coas shack hartmann aberrometer
Date Issued
01-01-2015
Author(s)
Abstract
In any optical system, optical aberrations of the imaging system affect the image quality. The human eye is also like an optical system which has optical aberrations influencing the quality of the retinal image. When pupil size exceeds 3 mm, ocular aberrations increase and play a major role on retinal image degradation. Pupil diameter is made constant in commercially available aberrometers by mathematically rescaling it. The aim of this study is to validate the pupil rescaling technique by using COAS (Complete Ophthalmic Analysis System)Shack Hartmann Aberrometer. Five subjects were recruited for this study. The measurements were taken over a moderately large pupil of 5mm in normal room illumination to allow for natural pupil dilation. The analyses diameter is fixed at 5 mm in COAS which means it rescales the aberration data to 5 mm if the pupil diameter recorded was more than 5 mm at the time of measurement. Ocular aberrations for natural and rescaled pupil sizes were analyzed. Estimation of ocular aberrations showed there was no statistical significance between natural pupil and rescaled pupil diameter.