Fast frame rate rodent cardiac x-ray imaging using scintillator lens coupled to CMOS camera
Date Issued
01-01-2017
Author(s)
Swathi Lakshmi, B.
Sai Varsha, M. K.N.
Kumar, Ashwin N.
Indian Institute of Technology, Madras
Indian Institute of Technology, Madras
Abstract
Micro-Computed Tomography (MCT) systems for small animal imaging plays a critical role for monitoring disease progression and therapy evaluation. In this work, an in-house built micro-CT system equipped with a X-ray scintillator lens coupled to a commercial CMOS camera was used to test the feasibility of its application to Digital Subtraction Angiography (DSA). Literature has reported such studies being done with clinical X-ray tubes that can be pulsed rapidly or with rotating gantry systems, thus increasing the cost and infrastructural requirements.The feasibility of DSA was evaluated by injected Iodinated contrast agent (ICA) through the tail vein of a mouse. Projection images of the heart were acquired pre and post contrast using the high frame rate X-ray detector and processing done to visualize transit of ICA through the heart.
Volume
10132