Options
ARTSENS - An image-free system for noninvasive evaluation of arterial compliance
Date Issued
31-10-2013
Author(s)
Indian Institute of Technology, Madras
Thomas, Emy Anu
Indian Institute of Technology, Madras
Suresh, S.
Abstract
Evaluation of arterial compliance is significant in cardiovascular diagnosis for early detection of coronary heart disease. We present ARTSENS, an image-free system for non-invasive evaluation of arterial compliance in-vivo. The system utilizes a single element ultrasound probe with intelligent measurement algorithms to ensure accurate evaluation of local arterial compliance without an image. The ability of the system to detect artery anatomy and measure compliance was verified by in-vivo measurements conducted on 106 subjects. The accuracy of compliance estimates were evaluated by comparison with a state of art imaging system. The measurements made using ARTSENS showed strong correlation with those made using the imaging system. The ability of ARTSENS to detect age-related trends in arterial compliance was also investigated. © 2013 IEEE.