Options
A Magnetic Plethysmograph Probe for Local Pulse Wave Velocity Measurement
Date Issued
01-10-2017
Author(s)
Nabeel, P. M.
Indian Institute of Technology, Madras
Indian Institute of Technology, Madras
Abstract
Objective: We present the design and experimental validation of an arterial compliance probe with dual magnetic plethysmograph (MPG) transducers for local pulse wave velocity (PWV) measurement. The MPG transducers (positioned at 23 mm distance apart) utilizes Hall-effect sensors and permanent magnets for arterial blood pulse detection. Methods: The MPG probe was initially validated on an arterial flow phantom using a reference method. Further, 20 normotensive subjects (14 males, age = 24 ± 3.5 years) were studied under two different physical conditions: 1) Physically relaxed condition, 2) Postexercise condition. Local PWV was measured from the left carotid artery using the MPG probe. Brachial blood pressure (BP) was measured to investigate the correlation of BP with local PWV. Results: The proposed MPG arterial compliance probe was capable of detecting high-fidelity blood pulse waveforms. Reliable local pulse transit time estimates were assessed by the developed measurement system. Beat-by-beat local PWV was measured from multiple subjects under different physical conditions. A profound increment was observed in the carotid local PWV for all subjects after exercise (average increment = 0.42 ± 0.22 m/s). Local PWV values and brachial BP parameters were significantly correlated (r ≥ 0.72), except for pulse pressure (r = 0.42). Conclusion: MPG arterial compliance probe for local PWV measurement was validated. Carotid local PWV measurement, its variations due to physical exercise and correlation with BP levels were examined during the in vivo study. Significance: A novel dual MPG probe for local PWV measurement and potential use in cuffless BP measurement.
Volume
11