Options
Analysis of a new polycentric above-knee prosthesis with a pneumatic swing phase control
Date Issued
01-01-1991
Author(s)
Patil, K. M.
Chakraborty, J. K.
Abstract
The four-bar linkage mechanism used by many designers for the knee joint of an above-knee prosthesis can provide knee rotation suitable for walking only. In Afro-Asian countries pepole are habituated to the squatting posture in their daily activities. A six-bar linkage knee-ankle mechanism above-knee prosthesis with a pneumatic swing phase control is developed and its analysis shows its suitability for producing a motion closely resembling that of a normal subject during walking and squatting. © 1991.